INTRODUCTION
This section describes performance and safety test procedures. Perform the
tests in this section upon completion of any repair work. If a monitor fails to
perform as specified in any test, repair the monitor, and repeat the safety and
performance test. Continue until the monitor passes all tests. You need not
remove the monitor cover to perform these tests.
Note:
Incorrect performance of the procedures described in this section may
affect monitor safety or accuracy. To ensure continued accuracy, you
must perform the procedures as described, using only the test
equipment specified in this section.
REQUIRED TEST EQUIPMENT
Table 7-1 lists required test equipment. The first item in the table is a special
test adapter you must fabricate (figure 7-1). Likewise, there are two fixtures
you must fabricate.
Table 7-1:
Required Test Equipment
Item
Description
.
Test adapter one
Fabricated by customer
Bio-tek 370 or equivalent safety | Must meet current Association for the
analyzer
Advancement of Medical
Instrumentation (AAMD) specifications
Digital multimeter
Fluke Type 87 or equivalent
Durasensor
Nellcor DS-100A
RCAL HI fixture
Fabricated by customer
RCALLO fixture
Fabricated by customer
Oxisensor
Nellcor D-25
Pulse oximeter tester
Nellcor PT-2500 or equivalent
BUILDING THE TEST ADAPTERS AND FIXTURES
You need the following parts to build test adapter one (figure 7-1):
One 9-pin male D-type connector
*
One connector hood
*
Wire
Here are the directions for building test adapter one:
L
Connect a conductor to each pin of the connector.
2..
Connect respective ends of each conductor together.
8.
Solder ends. .