Modifying and Testing the Oximeter
Once the functional tester is attached to the cable and the unit is turned on,
complete all of the tests in sequence, beginning with BPM, then %SpO
Modulation, and finally Light Level.
Overview
The SRC-MAX functional tester enables qualified technicians to functionally test
Nellcor OxiMax technology-based oximeters and OEM OxiMax technology-based
118
To Purchase, Visit
Avobus.com
Figure 83. SRC-MAX OxiMax Oximetry Tester
Table 21. SRC-MAX Callouts
1. DOC-10 Cable Connector
2. Infrared LED Drive Indicator
3. Pulse Rate Selection Button
4. Light Level Selection Button
10030881 Rev. B 12-2008
or call
1-800-674-3655
5. %SpO
Select Button
2
6. % Modulation Select Button
7. Battery Low Indicator
8. Red LED Drive Indicator
, then
2