Nellcor OXIMAX N-600X Series Service Manual page 90

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Managing the Data Port
3. Firmware Version—The next data field displays the firmware level (Version
2.0.4.0) and a firmware verification number (CRC: XXXX). Neither of these numbers
should change during normal operation.
Note:
The numbers may change if the oximeter is serviced and receives a
4. Alarm Limits—The last data field in the top line indicates the upper and the
lower alarm limits for SpO
lower alarm limit for is 85% and the upper alarm limit is 100%. Pulse Rate alarm
limits are 40 and 170 bpm. The SatSeconds alarm limit (100SAT-S) displays the
SatSeconds alarm setting. In this example, SatSeconds is set to 100.
5. Monitor Mode—The first field of the second line identifies the monitor (ADULT or
NEONATE) mode.
6. Response Mode—The second field of the second line identifies the SatSeconds
alarm limit and SpO
impact the SPD alarm behavior.
7. Data Column Headings—Actual column headings are in the last row of the
column heading line.
90
Figure 49. Location of Firmware Version in Column Headings
and for the pulse rate (PR). In the example above the
2
Figure 50. Location of Alarm Limits in Column Headings
Figure 51. Location of Monitor Mode in Column Headings
response (NORMAL or FAST) mode. The response mode may
2
Figure 52. Location of and Response Mode in Column Headings
Figure 53. Location of Data Column Headings
10030881 Rev. B 12-2008
firm
ware upgrade.
Service Manual