9. Press the SRC-MAX %SpO
The oximeter displays three dashes [ - - - ] and stabilizes at a value between 73
and 77 inclusive.
Figure 102. PLETH Display with SRC-MAX decrease in %SpO
a. Active audio alarm.
b. Flashing %SpO
c. BPM indication between 57 and 63 inclusive.
d. Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual amplitude may
vary, but can be used as a reference for low modulation/high light indications.
10. Press the SRC-MAX MODULATION selection button. The SRC-MAX MODULATION
LED lights. The oximeter pulse amplitude waveform increases in amplitude.
Figure 103. PLETH Display with SRC-MAX increase in modulation with high light condition
a. Active audio alarm.
b. Flashing %SpO
c. BPM indication between 57 and 63 inclusive.
d. Pulse waveform of approximately 1-inch P-T-P amplitude. Actual amplitude may vary,
but can be used as a reference for high modulation/high light indications.
11. Turn off the oximeter.
Service Manual
selection button. The SRC-MAX %SpO
2
indication between 73 and 77 inclusive.
2
indication between 73 and 77 inclusive.
2
10030881 Rev. B 12-2008
Performance Tests
75 LED lights.
2
of 75 with high light condition
2
129