Nellcor OXIMAX N-600X Series Service Manual page 131

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

All service function menus are accessible when the DOC-10 pulse oximetry cable is
disconnected from the oximeter. Disconnect the DOC-10 pulse oximetry cable from
the oximeter or disconnect the OxiMax pulse oximetry sensor from the DOC-10 pulse
oximetry cable.
Note:
Service function menus are only accessible from the Main Menu display.
Note:
With the DOC-10 pulse oximetry cable connected, only the PARAM and EXIT softkeys appear
on the screen.
To access service function menus:
1. Turn on the oximeter by pressing the ON/STANDBY button.
2. Wait for oximeter power-on self-test to complete.
3. Simultaneously press and hold the LIGHT softkey and the HELP/CONTRAST
button until the service softkeys display.
Navigational Softkeys
1. SELECT Softkey—SELECT the desired function.
2. NEXT Softkey—Access additional options for the Service Functions Menu by
pressing the NEXT softkey.
3. BACK Option—Return back to the SERVICE Main Menu.
4. EXIT Option—Exit the Service Menu to the Main Menu.
Main Service Menu
The following list can be used as a quick reference showing how to reach different
softkey functions. Items reached through the PARAM softkey can be accessed during
normal operation. Functions provided by the PRINT and NEXT softkeys cannot be
accessed when a pulse oximetry sensor is connected to the OxiMax N-600x pulse
oximeter. The Service Menu softkey options provide access to several submenus.
1. PARAM Menu—Reset and save parameters to factory defaults. Enable or disable
patient alarm events.
2. PRINT Menu—Print trend data recorded for up to 45 hours. Access a list of all
error log data recorded to memory. Retrieve or delete current trend data via an
Instat printout.
Service Manual
Figure 105. PARAM and PRINT Service Menu
10030881 Rev. B 12-2008
Performance Tests
131