Repairing the Oximeter
Battery Replacement
Removing the Old Battery
To remove the old battery:
1. Turn the oximeter off by pressing the ON/STANDBY button.
2. Disconnect the oximeter from the AC power source.
3. Follow the steps in Disassembling the Oximeter, page 167.
4. Remove the two screws from the battery bracket and lift the battery out of the
bottom case as shown.
5. Be sure to note the polarity of the leads. Use needle-nose pliers to disconnect the
leads from the battery.
Caution
The lead-acid battery is recyclable. Do not dispose of the battery by placing it in the regular
trash. Dispose of the battery in accordance with local guidelines or contact Nellcor's
Technical Services department to arrange for disposal.
170
Figure 127. Removing the Battery
10030881 Rev. B 12-2008
POX_10115_A
Service Manual