Low And Critical Battery Conditions - Nellcor OXIMAX N-600X Series Service Manual

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Troubleshooting

Low and Critical Battery Conditions

Critical
State
Battery
1
No
2
No
3
No
4
No
5
Not used
156
Table 32. Low and Critical Battery Conditions
Low
AC
Battery
Power
No
Yes
No
No
Yes
No
Yes
Yes
10030881 Rev. B 12-2008
Operation
SpO
-normal
2
AC/Battery charge LED-on
LOW BATTERY LED-off
LOW BATTERY message-off
Audible alarm-off
Error code-none
Effect of ALARM SILENCE key-normal
Shutdown-N/A
SpO
-normal
2
AC/Battery charge LED-off
LOW BATTERY LED-off
LOW BATTERY message-off
Audible alarm-off
Error code-none
Effect of ALARM SILENCE key-normal
Shutdown- N/A
SpO
-normal
2
AC/Battery charge LED-off
LOW BATTERY LED-on
LOW BATTERY message-on
Audible alarm-low priority
Error code-logged
Effect of ALARM SILENCE key-First press silences audio alarm,
second press cancels LOW BATTERY message (LED) stays on until
Low Battery Condition is corrected.
Shutdown-Imminent
SpO
-normal
2
AC/Battery charge LED-on
LOW BATTERY LED-on
LOW BATTERY message-off
Audible alarm-off
Error code-logged
Effect of ALARM SILENCE key-N/A (LED stays on)
Shutdown-N/A