Alarm Speaker Replacement
Removing the Alarm Speaker
WARNING
If replacing the speaker because of speaker failure, please contact Nellcor Technical
Services or your local Nellcor representative.
Note:
Contact Nellcor Technical Services to discuss the return or disposal of the original speaker
assembly.
To remove the alarm speaker:
1. Turn the oximeter off by pressing the ON/STANDBY button.
2. Disconnect the oximeter from the AC power source and follow the steps outlined
in Disassembling the Oximeter, page 167.
3. Disconnect the speaker wire harness from J9 (1) on the Main Board PCB as shown.
4. Pull the holding clip (2) back from the speaker and lift the speaker out of the top
case.
Service Manual
To Purchase, Visit
Avobus.com
Figure 133. Removing the Alarm Speaker
1
10030881 Rev. B 12-2008