6. Press the SELECT softkey to select PROTOCOL.
7. Set the PROTOCOL to ASCII for text printing using the ADJUST UP or ADJUST
DOWN key below the ALARM SILENCE key on the oximeter panel if the protocol is
different. For print standards, refer to Data Port Connectivity, page 79.
8. Press the EXIT softkey.
9. Enter the Service Menu by simultaneously pressing and holding the LIGHT softkey
and the HELP/CONTRAST button until the service softkeys display.
10.Press the PRINT softkey to reach the options for TREND, ERRLOG, INSTAT, and INFO
printouts.
11.Select the appropriate printout with the corresponding softkey. For more
information about each option, refer to PRINT Softkey Menu, page 132.
Note:
Once trend printing begins, printing can only be aborted by turning off the N-600x or the
printer.
Service Manual
Figure 42. Serial Port Setup Screen, Protocol Selection
Figure 43. PARAM and PRINT Service Menu
Figure 44. PRINT Service Menu Options
10030881 Rev. B 12-2008
Managing the Data Port
87