Replacing the Battery
To replace the old battery:
1. Follow the steps outlined in Removing the Old Battery, page 170.
2. Connect the leads to the battery. The red wire connects to the positive terminal,
and the black wire connects to the negative terminal.
3. Insert the new battery into the bottom case with the negative terminal towards
the outside of the oximeter. Install the bracket and grounding lead with the two
screws.
4. Follow the steps outlined in Reassembling the Oximeter, page 169.
5. Power the oximeter on to verify proper operation.