Table 7. Electromagnetic Immunity Testing
IEC 60601-1-2
Immunity Test
Test Level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2:
2001
Electric fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4:
± 1 kV for input/
1995 + A1: 2000 +
output lines
A2: 2001
Surge
± 1 kV differential
mode
IEC 61000-4-5:
2005
± 2 kV common
mode
Voltage dips,
<5% UT
short
(>95% dip in UT)
interruptions and
for 0.5 cycle
voltage variations
on power supply
40% UT
IEC 61000-4-11:
(60% dip in UT)
2004
for five cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(95% dip in UT)
for five seconds
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8:
2001
10030881 Rev. B 12-2008
or call
1-800-674-3655
Compliance
Electromagnetic
Level
Environment Guidance
± 6 kV contact
Floor should be wood,
concrete, or ceramic tile. If
± 8 kV air
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
± 2 kV for power
Mains power quality should
supply lines
be that of a typical
commercial and/or hospital
± 1 kV for input/
environment
output lines
± 1 kV differential
Mains power quality should
mode
be that of a typical
commercial and/or hospital
± 2 kV common
environment
mode
<5% UT
Mains power quality should
be that of a typical
(>95% dip in UT)
commercial and/or hospital
for 0.5 cycle
environment.
40% UT
If the user of the oximeter
requires continued
(60% dip in UT)
operation during power
for five cycles
mains interruption, it is
70% UT
recommended that the
oximeter be powered from
(30% dip in UT)
an uninterruptible power
for 25 cycles
supply or battery.
<5% UT
Note: UT is the AC main's
(95% dip in UT)
voltage prior to
for five seconds
application of the test
level.
3 A/m
If image distortion occurs, it
may be necessary to position
the oximeter further from
the sources of power
frequency magnetic fields or
to install magnetic shielding.
The power frequency
magnetic field should be
measured in the intended
installation location to
assure that it is sufficiently
low.