Sensor And Cable Compliance - Nellcor OXIMAX N-600X Series Service Manual

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Immunity Test
Conducted RF
IEC
61000-4-6: 2006
Radiated RF
IEC 61000-4-3: 2006
Rated Maximum
Output Power of
Transmitter in Watts
0.01
0.10
1.00
10.00
100.00
For transmitters rated at a maximum output power not listed above, the separation
distance can be estimated using the equation in the corresponding column, where P
is the maximum output [power rating of the transmitter in watts (W)] according to
the transmitter manufacturer.
Note:
Portable and mobile RF communications equipment should be used no closer to any part of
the OxiMax N-600x pulse oximeter, including cables, than the recommended separation
distance calculated from the equation appropriate for the frequency of the transmitter.

Sensor and Cable Compliance

WARNING
The use of accessories, OxiMax sensors, and cables other than those specified may
result in inaccurate readings of the OxiMax N-600x pulse oximeter and increased
emission of the oximeter.
Service Manual
Table 8. Recommended Separation Distances
IEC 60601-1-2 Test
Level
Compliance Level
Frequency of
Transmitter
3 Vrms
3 Vrms
150 kHz
80 MHz
3 V/m
3 V/m
80 MHz
800 MHz
3 V/m
3 V/m
800 MHz
2.5 GHz
Separation
Distance in Meters
Distance in Meters
0.12
0.12
0.38
0.38
1.20
1.20
3.80
3.80
12.00
12.00
10030881 Rev. B 12-2008
Manufacturer's Declaration
Electromagnetic
Environment
Guidance
Equation for
Separation
Distance
distance = 1.2
distance = 1.2
distance = 2.3
Separation
Separation
Distance in Meters
0.23
0.73
2.30
7.30
23.00
35