Performance Tests
Alarms and Alarm Silence
Figure 72. SpO
Sensor Port
2
To adjust the alarms and alarm silence options:
1. Connect the DOC-10 pulse oximetry cable to the oximeter's SpO
Sensor Port.
2
2. Connect the OxiMax DS-100A sensor to the DOC-10 cable, then to a finger.
3. Press the ON/STANDBY button to turn the oximeter on.
4. Press SETUP and then VIEW.
5. Press the PLETH softkey. Verify the %SpO
and BPM indicate the SpO
and pulse rate.
2
2
6. Press the LIMITS softkey.
7. Press the SELECT softkey to select %SpO
Lower Alarm Limit.
2
8. Press the ADJUST UP button until the %SpO
Lower Alarm Limit indicates 99.
2
Figure 73. Adult %SpO
Lower Alarm Limit of 99
2
9. Press the SELECT softkey four consecutive times to select Pulse Rate Lower Alarm
Limit.
10. Press the ADJUST UP button until the Pulse Rate Lower Alarm Limit indicates 160.
Figure 74. Adult Lower BPM Alarm Limit of 160
10030881 Rev. B 12-2008
111