Nellcor OXIMAX N-600X Series Service Manual page 85

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Note:
To save the baud rate setup for the oximeter as institutional default settings, See Adjusting
the Factory Default Settings, page 72. Use care when performing this procedure, since all
settings will be saved as institutional default settings.
To use the communication protocol:
Use the COMM softkey to select the appropriate communication protocol supported
for use with the data port.
ASCII used for printouts and Oxinet III output
CLINICAL intended for Nellcor use only
PHILIPS interfaces the OxiMax N-600x pulse oximeter with a Philips monitor
SPDout used for faster communication baud rates
Note:
Setting the communication protocol automatically sets the communication baud rate.
To change the communication protocol:
1. Turn on the oximeter by pressing the ON/STANDBY button.
2. Press the SETUP softkey.
3. Press the NEXT softkey twice.
4. Press the COMM softkey.
5. Press the SELECT softkey.
6. Use the ADJUST UP and ADJUST DOWN buttons to select the desired protocol.
7. Press the EXIT softkey set the protocol. The protocol setting is in effect until the
oximeter is powered off.
Note:
To save the protocol setup for the oximeter as an institutional default setting, see Adjusting
the Factory Default Settings, page 72. Use care when performing this procedure, since all
settings will be saved as institutional default settings.
Philips Communications
WARNING
Do not silence or decrease the volume of the OxiMax N-600x pulse oximeter audible
alarm if patient safety could be compromised.
Caution
Alarms still originate from the OxiMax N-600x pulse oximeter.
The OxiMax N-600x pulse oximeter sends SpO
the Philips monitor. The Philips monitor requires a Philips VueLink™ Aux Plus B
interface module (A05 option) to interface with the OxiMax N-600x pulse oximeter.
Service Manual
To Purchase, Visit
Avobus.com
10030881 Rev. B 12-2008
or call
1-800-674-3655
Configuring the Data Port Interface
, pulse rate, and alarm status data to
2
85