Display PCB Replacement
Removing the Display PCB
WARNING
The LCD panel contains toxic chemicals. Do not touch broken LCD panels. Physical
contact with a broken LCD panel can result in transmission or ingestion of toxic
substances.
To remove the Display PCB:
1. Turn the oximeter off by pressing the ON/STANDBY button.
2. Disconnect the oximeter from the AC power source and follow the steps outlined
in Disassembling the Oximeter, page 167.
3. Disconnect the Cold Cathode Fluorescent Light (CCFL) harness from J4 (6) of the
Main Board PCB as shown.
Service Manual
Figure 131. Removing the Display PCB
2
1
Table 39. Opening Oximeter Casting Figure Callouts
No
1
Speaker wires to J9 Connector on Main PCB
10030881 Rev. B 12-2008
Display PCB Replacement
3
4
5
6
7
Description
POX_10041_A
177