Replacing the Power Entry Module

To replace the power entry module:
1. Reconnect the three power supply leads as indicated in Removing the Power Entry
Module, page 172.
2. Install the PEM in the bottom case with the fuse drawer facing down. A tab in the
bottom case holds the PEM in place.
3. Insert the bottom wing of the PEM between the tab and the internal edge of the
sidewall of the bottom case.
4. Push the PEM down and towards the outside of the oximeter until it clicks into
place.
5. Follow the steps in Reassembling the Oximeter, page 169.
Service Manual
Power Entry Module (PEM) Replacement
10030881 Rev. B 12-2008