Modifying and Testing the Oximeter
8. Press the ADJUST UP button repeatedly until the SPD alarm threshold cycles from 2 to 3, from
3 to OFF, and back to 1.
9. Press the SELECT softkey three times. Confirm the Pulse Rate Delay feature is selected.
10. Confirm the Pulse Rate Delay is set to OFF.
11. Press the ADJUST UP button repeatedly until the Pulse Rate Delay alarm threshold cycles from
OFF through 5, 10, and back to OFF.
12. Press the ON/STANDBY button to turn the oximeter off.
13. Press the ON/STANDBY button to turn the oximeter back on.
14. Press the LIMITS softkey.
15. Verify SatSeconds indicates an alarm limit of "100, " Saturation Pattern Detection (SPD)
indicates a sensitivity setting of "1, " and Pulse Rate Delay is OFF.
16. Press the ON/STANDBY button to turn the oximeter off.

Operational Setup

Operational Setup procedures confirm settings for and allow configuration of the
listed parameters.
Battery Charge, page 106
Alarms and Alarm Silence, page 111
Alarm Volume Control, page 113
Pulse Tone Volume Control, page 114
Using the Analog Output, page 114
110
Figure 70. Testing Saturation Pattern Detection Thresholds
Figure 71. Testing Pulse Rate Delay Thresholds
10030881 Rev. B 12-2008