Accessing Data Using The Comm Port And A Computer - Nellcor OXIMAX N-600X Series Service Manual

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

5. Press the ADJUST UP or ADJUST DOWN softkeys to select the desired protocol.
Protocol options for the OxiMax N-600x pulse oximeter differ from those of the
oximeter with the SPD feature activated. The factory default protocol is ASCII. See
Communication Protocol Compatibility and Output, page 83.
Communication
Protocols
ASCII
CLINICAL
GRAPH (Graphics)
OXINET
PHILIPS
SPACELBS (Spacelabs)
MARQ (GE Marquette)
DATEX
(Datex-Ohmeda)
SPDout
1
Not provided in real-time output
6. Press the SELECT softkey.
Note:
When SPD is enabled, the Clinical and SPDout protocols include SPD information. Results
cannot be predicted for host systems without upgrades to support the extra SPD
information. See
83.
7. Press the ADJUST UP or ADJUST DOWN softkeys below the ALARM SILENCE key
on the oximeter panel to select the desired baud rate, should it differ from the
default for that particular protocol.
8. Press the EXIT softkey.

Accessing Data Using the Comm Port and a Computer

Capture the data displayed on the computer screen and use as a text file or in a
spreadsheet.
To access data using a computer:
1. Open a terminal program, such as Hyper Terminal.
Service Manual
Table 16. Communication Protocol Compatibility and Output
N-600x with
Firmware up to 2.0
Default Baud Rate
19200
19200
N/A
Table 16: Communication Protocol Compatibility and Output
10030881 Rev. B 12-2008
N-600x with Firmware 2.0 and Above
Default Baud Rate
9600
9600
19200
9600
N/A
9600
N/A
19200
Use only 19200
9600
9600
9600
N/A
2400
N/A
Default
115200
Available:
19200, 57600, 115200
Data Port Connectivity
Output includes SPD
data fields
1
Yes
Yes
N/A
N/A
No
N/A
N/A
N/A
Yes
, on page
83