Oximetry Considerations; Patient Conditions; Oximax™ Pulse Oximetry Sensor Performance Considerations; Safety Information - Nellcor OXIMAX N-600X Series Service Manual

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Performance Considerations

Oximetry Considerations

Pulse Rates
The oximeter only displays pulse rates between 20 and 250 bpm. Detected pulse
rates above 250 bpm are displayed as 250. Detected pulse rates below 20 are
displayed as a zero (0).
Saturation
The oximeter displays saturation levels between 1% and 100%.

Patient Conditions

Dysfunctional Hemoglobins
Dysfunctional hemoglobins such as carboxyhemoglobin, methemoglobin, and
sulphemoglobin are unable to carry oxygen. SpO
however, a patient may be hypoxic because less hemoglobin is available to carry
oxygen. Further assessment beyond pulse oximetry is recommended.
Anemia
Anemia causes decreased arterial oxygen content. Although SpO
appear normal, an anemic patient may be hypoxic. Correcting anemia can improve
arterial oxygen content. The oximeter may fail to provide an SpO
hemoglobin levels fall below 5 gm/dl.
OxiMax™ Pulse Oximetry Sensor Performance Considerations

Safety Information

WARNING
Pulse oximetry readings and pulse signal can be affected by certain ambient
conditions, sensor application errors, and certain patient conditions.
WARNING
Tissue damage can be caused by incorrect application or inappropriate duration of use
of an OxiMax pulse oximetry sensor. Inspect the sensor site as directed in the
Directions for Use.
WARNING
Use only Nellcor-approved OxiMax pulse oximetry sensors and pulse oximetry cables
when connecting to the OxiMax sensor connector. Connecting any other cable or
sensor influences accuracy of sensor data, which may lead to adverse results.
WARNING
Failure to cover the OxiMax pulse oximetry sensor site with opaque material in high
ambient light conditions may result in inaccurate measurements.
98
10030881 Rev. B 12-2008
readings may appear normal;
2
readings may
2
reading if
2
Service Manual

Hide quick links:

Advertisement

Table of Contents