Nellcor OXIMAX N-600X Series Service Manual page 95

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

holds
that point for 15 seconds. It increases to 1.0 VDC and
seconds. The third part of the calibration signal is a stair step signal. The stair step
signal, begins at 0.0 VDC and increases up to 1.0 VDC in 0.1 VDC increments, each
held for one (1) second. Select the 0.0 VDC, 1.0 VDC, or stair step signal individually,
using the ANALOG softkey menu option.
The OxiMax N-600x pulse oximeter data port provides analog voltage outputs
between pins 6, 13, 14, and ground (pin 10), which can be used to calibrate oximeters
such as a chart recorder. The voltage represents a specific measured parameter's
current value. The voltage differential varies proportionally from 0.0 to +1.0 VDC as
the pin's parameter varies over its full range of values. For example, as the current
value of SpO
from 0.0 to +1.0 VDC. A voltage of 0.94 volts indicates a current SpO
To set analog voltage output
1. Press the SETUP softkey.
2. Press the NEXT softkey three times.
3. Press the ANALOG softkey.
a. VOLT Options—Selecting the 0 VOLT or 1 VOLT softkey causes that voltage to appear
at pins 6, 13, or 14 as referenced to ground pins 5 and 10.
b. STEP Option—Selecting the STEP softkey causes the voltage to increase from 0.0 to
1.0 volts at 1/10th-volt increments, with each step lasting at least one second.
4. Press the BACK softkey.
Service Manual
Figure 59. Analog Pinouts
Pin
Parameter
6
Saturation
13
Pulse Rate
14
Waveform
varies from 0% to 100%, the voltage from pin 6 to ground (pin 10) varies
2
Figure 60. Data Port Pin Layout
Figure 61. Analog Voltage Screen
10030881 Rev. B 12-2008
Calculating the Analog Voltage Output
VDC
Parameter Range
0 - 100%
0 - 250 bpm
0 - 254 PAU
holds
that value for 15
value of 94.
2
95