Nellcor OXIMAX N-600X Series Service Manual page 67

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

TREND Menu
Choose the type of trend data to view by selecting trend data from the oximeter
(MONITR) or trend data from the pulse oximetry sensor (SENSOR) in the Trend menu.
1. MONITR Menu—Isolate oxygenation (SpO
them both together (DUAL) for a specified length of time. Establish trend data
parameters for specific time segments and set minimum and maximum trend
values.
a. VIEW Menu—Isolate oxygenation (SpO
b. ZOOM Menu—Access to the TIME option, SCALE option, AUTO option, and BACK
Note:
The ZOOM menu will not allow access to the options listed if it is in AMP mode. SCALE and
AUTO options do not appear as options without trend data history to review.
Service Manual
To Purchase, Visit
Avobus.com
both together (DUAL). Access the HIST and Amplitude submenus by selecting the
NEXT Option.
- DUAL softkey—View oxygenation
for a specified length of time.
- SpO
softkey—Isolate oxygenation
2
time.
- PULSE softkey—Isolate pulse (PULSE) trend data for a specified length of time.
- HIST softkey—Choose to delete or keep current trend data history. To delete,
select YES. To keep, select NO and return to the HIST menu, go BACK, or EXIT to
the Main Menu.
- AMP softkey—Set the trend data view to "Pulse Amplitude Units" (PAU) using
the AMP (amplitude) Menu. This is an arbitrary unit of measure gauging the
distance from the peak to the valley of a pulse waveform. The trend data when
set to PAU can register amplitudes up to 200 PAU.
option if the trend view is set in DUAL, SpO
- TIME softkey—Select from hours to minutes to seconds for viewing a specific
time segment of trend data. Cycle through the options by pressing the TIME
softkey. The options proceed in the following order: 48h, 36h, 24h, 12h, 8h, 4h,
2h, 1h, 30m, 15m, 40s, or 20s.
- SCALE softkey—Select the maximum and minimum values for the SAT or
PULSE trend graph. Default SAT trend values are 10 to 100. Default PULSE trend
values are 5 to 250. Cycle through the options by pressing the Select softkey.
The options proceed in the following order: ±5, ±10, ±15, ±20, ±25, ±30, ±35,
±40 and ±50 units variance.
- AUTO softkey—Obtain the maximum and minimum rounded values based on
all graphed trend data.
- DELETE softkey—Choose to delete or keep current trend data. To delete,
select YES. To keep, select NO and return to the previous menu.
or call
1-800-674-3655
) or pulse (PULSE) trend data or view
2
) or pulse (PULSE) trend data or view them
2
(
SpO
) and pulse (BPM) trend data together
2
(
SpO
) trend data for a specified length of
2
, or PULSE mode.
2
10030881 Rev. B 12-2008
Using Oximeter Softkey Menus
67