Advertisement

5. Press the SRC-MAX LIGHT LEVEL selection button. The SRC-MAX LIGHT
LEVEL
and stabilizes at the same amplitude.
Note:
Flat-lining is the only indication of a light change at the measurement site. If the oximeter
recovers and displays normally, this is an indication of proper operation with light changes.
a. Active audio alarm.
b. Flashing %SpO
c. BPM indication between 57 and 63 inclusive.
d. Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual amplitude may
6. Press the SRC-MAX PULSE RATE selection button. The SRC-MAX PULSE RATE
200 LED lights. The oximeter registers the BPM increases and stabilizes at a value
between 197 and 203 inclusive.
a. Active audio alarm.
b. Flashing %SpO
c. Flashing BPM indication between 197 and 203 inclusive.
d. Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual amplitude may
Service Manual
To Purchase, Visit
Avobus.com
LED lights. The oximeter pulse amplitude waveform initially flatlines
Figure 98. PLETH Display with SRC-MAX increase in light
indication between 73 and 77 inclusive.
2
vary, but can be used as a reference for low pulse amplitude/high light patients.
Figure 99. PLETH Display with SRC-MAX increase to 200 BPM with high light condition
indication between 73 and 77 inclusive.
2
vary, but can be used as a reference for low modulation/high light indications.
or call
1-800-674-3655
10030881 Rev. B 12-2008
Performance Tests
127