f. Use the SELECT softkey to toggle between ALLOW OFF, OFF REMINDER, or ALLOW
SAT-S options. Use the ADJUST UP or ADJUST DOWN button to change selected
parameter.
- ALLOW OFF—Enable or disable the AUDIBLE ALARM OFF option. Press the
- OFF REMINDER—Enable or disable the reminder tone for when the alarm off
- SPD Setup—This option is only available after activation of the SPD feature by
g. Press the BACK softkey.
3. BATTRY Option—Reserved for Nellcor use only. Do not use.
Service Manual
ADJUST UP or ADJUST DOWN button to toggle between YES and NO. If YES is
selected, the option of selecting AUDIBLE ALARM OFF as an alarm silence
duration choice becomes available. If NO is selected, no additional option
becomes available.
option has been used. If the audible alarm is set to OFF, a reminder tone can be
sounded every three minutes to notify the user of this condition. The ADJUST
UP and ADJUST DOWN buttons can be used to change the choice from YES to
NO. Selecting YES enables the Reminder. Selecting NO disables the Reminder
when the audible alarm is set to OFF.
a qualified service technician. Enable or disable the ALLOW SPD option and
select from the AUDIO ALERT options. Use the ALLOW SPD option to enable or
disable OxiMax SPD™ Alert saturation pattern detection. If SPD is enabled, the
SPD feature is available for adults in clinical settings and also automatically
enables Sat-Seconds, setting the Sat-Second value at 100. Use the AUDIO
ALERT option to select AUD VIS for both audio and visual SPD alarms (tonal
alarm and flashing TREND key), or to select VIS for visual only SPD alarms
(flashing TREND key only).
Figure 110. SPD Service Menu Options
Figure 111. BATTRY Service Menu Options
10030881 Rev. B 12-2008
Performance Tests
135