oximeters. The following table provides a brief description of each test. Use the
PLETH view for all functional testing.
Note:
The SRC-MAX selectable indicator LEDs may extinguish if there is a delay in proceeding
through the above tests. This is normal operation in order to increase the battery time.
Note:
Pressing a button on the SRC-MAX during the test procedures may be requested, changing a
certain parameter. If the SRC-MAX LEDs are not lit, press the button twice. Pressing the
button once causes the indicators to relight and pressing twice initiates the change.
BPM Test
1. With the oximeter turned off, connect the DOC-10 pulse oximetry cable to the
sensor port.
2. Connect the SRC-MAX tester to the other end of the DOC-10 cable.
3. Turn on the oximeter by pressing the ON/STANDBY button. Wait for the oximeter
to complete its self-check.
Service Manual
Table 22. Functional Tests Options
Tests
BPM Test
The test procedure simulates an OxiMax pulse
oximetry sensor attached to a patient indicating a
pulse rate of 60 BPM and 200 BPM.
2
%SpO
Test
The test procedure simulates an OxiMax pulse
oximetry sensor attached to a patient indicating a 75%
blood oxygen saturation and 90% blood oxygen
saturation.
Modulation Level Test
The test procedure simulates an OxiMax pulse
oximetry sensor attached to a patient indicating low
and high pulse strength.
Light Level Test
The test procedure simulates an OxiMax pulse
oximetry sensor attached to a patient indicating low
and high light level passing through the patient at the
sensor site.
10030881 Rev. B 12-2008
Descriptions
Performance Tests
119