6. Press the SRC-MAX PULSE RATE selection button. The SRC-MAX PULSE RATE
60 LED lights. The oximeter registers BPM decreases and stabilizes to a value
between 67 to 73 BPM inclusive.
a. Active audio alarm.
b. Flashing %SpO
c. BPM indication between 57 and 63 inclusive.
d. Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual amplitude may
7. Turn off the oximeter.
SpO
1. With the oximeter turned off, connect the DOC-10 pulse oximetry cable to the
sensor port.
2. Connect the SRC-MAX tester to the other end of the DOC-10 cable.
3. Turn on the oximeter by pressing the ON/STANDBY button. Wait for the oximeter
to complete its self-check.
4. Select the PLETH view by pressing the following softkeys in sequence: SETUP,
VIEW, and then PLETH to view the pleth display.
a. Active audio alarm.
b. Flashing %SpO
c. BPM indication between 57 and 63 inclusive.
d. Pulse waveform of approximately 1/2-inch peak to peak (P-T-P) amplitude. Actual
Service Manual
To Purchase, Visit
Avobus.com
Figure 86. PLETH Display with SRC-MAX decrease to 60 BPM
indication between 73 and 77 inclusive.
2
vary but will be a reference for low pulse amplitude/low light patients.
Test
2
Figure 87. PLETH Display with SRC-MAX tester-generated waveform
indication between 73 and 77 inclusive.
2
amplitude may vary but will be a reference for low pulse amplitude/low light patients.
or call
1-800-674-3655
10030881 Rev. B 12-2008
Performance Tests
121