Chapter 3. Product Specifications; Overview; Physical Characteristics; Electrical Requirements - Nellcor OXIMAX N-600X Series Service Manual

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

3 Product Specifications

Overview

This section contains OxiMax N-600x™ pulse oximeter physical and operational
specifications. Ensure all product requirements are met prior to installation of the
oximeter.

Physical Characteristics

Weight
Dimensions

Electrical Requirements

Power

Power Requirements
Input Frequency
Fuses

Battery

Note:
The battery provides approximately seven hours of battery life when new and fully-charged
with no alarms, no serial data, no analog output, no nurse call output, with backlight on
while using a pulse simulator set for 200 bpm, high light and low modulation.
Type
Voltage
Recharge
Shelf Life
Compliance
Service Manual
5.8 lbs. (2.6 kg)
3.3 in. x 10.4 in. x 6.8 in. (8.4 cm x 26.4 cm x 17.3 cm)
Rated at 100 to 120 volts AC (nominal 120 VAC) or
220 to 240 volts AC (nominal 230 VAC),
30 VA
50/60 Hz
Slow-blow 0.5 amp, 250 volts, IEC (5 x 20 mm)
Quantity: 2 external
Lead acid
6 Volts DC
8 hours with oximeter turned off
12 hours with oximeter turned on
Four months, if oximeter runs on new, fully-charged battery
After four months storage, units run 33% of stated battery life
91/157/EEC
10030881 Rev. B 12-2008