Chapter 11. Modifying And Testing The Oximeter; Overview; Setting Institutional Defaults (Sample) - Nellcor OXIMAX N-600X Series Service Manual

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

11 Modifying and Testing the Oximeter

Overview

This section provides information to trained service technicians on setting institutional
defaults, verifying OxiMax N-600x™ pulse oximeter performance, and accessing service
functions.
WARNING
Only qualified service personnel should open the oximeter housing, remove and
replace components, or make adjustments. If your medical facility does not have a
qualified service technician, please contact Nellcor's Technical Services or your local
Nellcor representative.

Setting Institutional Defaults (Sample)

Change power-on default values to institutional power-on default values using the
oximeter's service mode. Set temporary desired limits in normal operation mode.
Alarms
Alarm Volume
ALARM SILENCE Duration
Pulse Beep Volume
Pulse Rate Upper Alarm Limit
Pulse Rate Lower Alarm Limit
Pulse Rate Delay Feature
%SpO
%SpO
Allow SatSeconds
SatSeconds™ Alarm Management Feature
OxiMax SPD™ Alert (SPD) Feature
SPD Alarm Type
Sensor Event Record Type
Service Manual
To Purchase, Visit
Avobus.com
Table 19. Institutional Default Settings
Parameter
Upper Alarm Limit
2
Lower Alarm Limit
2
or call
1-800-674-3655
Allow Off - Yes/No
1 to 10 (min. 45dB, max > 85dB)
30, 60, 90, 120 seconds (Service Mode settings only)
0 to 10
Lower Limit plus 1 to 250 BPM
20 BPM to Upper Limit minus 1
OFF, 5, 10 seconds
Lower Limit plus 1 to 100%
Upper Limit minus 1 to 85%
Yes/No
OFF, 10, 25, 50, 100 SatSeconds
OFF, 1, 2, 3
Audible and Visual, Visual Only
SpO
, SpO
+ BPM, Default
2
2
10030881 Rev. B 12-2008
Possible Setting
103