P e r f o r m a n c e C o n s i d e r a t i o n s
Performance Verification
N-595 Monitor Performance
N-595
WARNING: Pulse oximetry readings and pulse signals can be
affected by certain ambient environmental conditions, O
sensor application errors, and certain patient conditions. See the
appropriate sections of the manual for specific safety
information:
•
Safety Information on page 1
•
OXIMAX Sensors and Accessories on page 99
•
Performance Considerations on page 111
The performance of the N-595 can be verified by following the
procedures outlined in the Performance Verification section of the
N-595 service manual. Qualified service personnel should perform
these procedures before using the monitor for the first time in a
clinical setting.
Considerations
Certain patient conditions can affect the measurements of the N-595
and cause the loss of the pulse signal.
Inaccurate measurements can be caused by:
•
prolonged and/or excessive patient movement
•
venous pulsations
M
XI
AX
111