Performance Verification; Overview; Required Equipment - Nellcor OxiMax N-600x Service Manual

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

P e r f o r m a n c e Ve r i f i c a t i o n

Overview

Required Equipment

O
M
N-600x Service Manual
XI
AX
This section discusses the tests used to verify performance following repairs or
during routine maintenance. All tests can be performed without removing the
N-600x cover. All tests except the battery charge and battery performance tests
must be performed as the last operation before the N-600x is returned to the
Caregiver.
If the N-600x fails to perform as specified in any test, repairs must be made to
correct the problem before the monitor is returned.
Table 1: Equipment and Descriptions
Equipment
Digital Multimeter (DMM)
®
Durasensor
O
M
Oxygen
XI
AX
Sensor
O
M
Oxygen Sensor
XI
AX
Safety Analyzer
Pulse Oximetry Cable
Data Interface Cable
Stop Watch
Nellcor model SRC-MAX Tester

Performance Verification

Description
Fluke Model 87 or equivalent
DS-100A sensor
MAX-A sensor
Must meet current AAMI ESI/1993 & IEC
60601-1/1998 specifications
DOC-10 cable
EIA-232 cable (optional)
Manual or electronic
Provides testing for DigiCal compatible
Monitors
15