Managing the Data Port

The RS-232 hardware interface cable has a DB-15 connector for the OxiMax N-600x
pulse oximeter and the applicable connector for the Philips monitor. For this
interface, Nellcor recommends Nellcor cable part number 902256.
A blank screen on the Philips monitor indicates corrupt data. The Philips monitor
detects corrupt data in less than 100 milliseconds. In the Philips mode of operation,
the OxiMax N-600x pulse oximeter automatically sets the interface baud rate to
19200 bits per second (bps).
Managing the Data Port

Data Printout Information

Printing Oximeter Trend Data
Trend information (oximeter and in-sensor event history) can be sent to a personal
computer or to a serial printer.
Note:
The protocol settings must be set to ASCII MODE for printing text data.
To print trend data
1. With the oximeter in the normal monitoring mode, connect the serial printer to
the oximeter's data port connector, using Nellcor printer cable part number
036341.
2. Turn on the printer.
3. Press the SETUP softkey on the monitor, and then press the NEXT softkey twice.
4. Press the COMM softkey.
5. Set the BAUD rate to the appropriate number using the ADJUST UP or ADUST
DOWN key.
86
Figure 41. Data Port Location
10030881 Rev. B 12-2008
Service Manual