Nellcor OXIMAX N-600X Series Service Manual page 143

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Message
DATA TYPE
SpO
+BPM
2
DEFAULTS
LOST
DELETE
TRENDS?
HELP
SPEAKER
FAILURE
INVALID
SILENCE
DURATION
INVALID
%SpO
LIMIT
2
LOW BATTERY
PRIMARY
SPEAKER
FAILURE
Service Manual
Table 24. Common Prompts and Error Messages
Time-
Exit on
out in
Exit on
Alarm
secs
Alarm?
Silence?
5
No
Yes
None
No
No
10
Yes
Yes
None
No
No
3
No
Yes
3
No
Yes
None
No
Yes
None
No
No
10030881 Rev. B 12-2008
Prompts, Error Messages, and Error Codes
Displayed
When a blank event
Clears on time-out, sensor
sensor is connected
disconnect, or pressing the
to a oximeter with
ALARM SILENCE key,
event data type set to
whichever comes first.
SpO
+BPM.
2
If the oximeter
Leave the factory default
detects that
settings, have a qualified
power-on settings
service technician change
have been lost.
institutional defaults, or
change temporary limit
settings.
When attempting to
Respond to the prompt,
delete trend data
electing to delete trends by
from memory by
selecting YES or to retain by
pressing the DELETE
selecting NO.
softkey.
After pressing the
Notify a qualified service
HELP softkey
technician. Once the oximeter
provided after a
powers down, it will not
Primary Speaker
power on again until the
Failure
technician replaces the failing
speaker.
When attempting to
If attempting to save defaults,
save current settings
turn ALARM SILENCE duration
as power-on defaults,
to a value, rather than OFF.
even though the
ALARM SILENCE
duration is OFF.
After attempting to
Set power-on %SpO
set the default %SpO
alarm limit default at or above
2
lower alarm limit
85.
below 85.
When the oximeter is
Connect to AC power to fully
on battery power and
charge or temporarily
the battery charge is
acknowledge by pressing the
low.
ALARM SILENCE key.
Fatal error caused by
Press the HELP softkey to
a hardware failure.
review the HELP SPEAKER
Once this error
FAILURE message. Notify a
displays, the only
qualified service technician.
option available is to
Once the oximeter powers
press the HELP
down, it will not power on
softkey, which
again until the technician
produces the HELP
replaces the failing speaker.
SPEAKER FAILURE
message.
Resolution
lower
2
143