Modifying and Testing the Oximeter
Sensor Adjust Enable
Nurse Call Priority RS-232
Data Port Mode
Available selections depend on firmware version.
Data Port Baud Rate
To set institutional defaults:
1. Disconnect the OxiMax pulse oximetry sensor from oximeter.
Note:
If the OxiMax pulse oximetry sensor is not disconnected, the only softkeys on the screen will
be PARAM and EXIT.
2. Set desired parameters to the institutional values. Refer to Adjusting the Factory
Default Settings, page 72, for parameter values.
3. Simultaneously press the LIGHT softkey and the HELP/CONTRAST button until
the softkey headings appear.
4. Press the PARAM softkey.
5. Press the SAVE softkey.
104
Table 19. Institutional Default Settings
Parameter
Figure 62. Service Main Menu
Figure 63. SAVE Option
10030881 Rev. B 12-2008
Possible Setting
Yes/No
Normally high, normally low
ASCII, CLINICAL, PHILIPS,SPDout
2400, 9600, 19200
Service Manual