Managing the Data Port
At the end of the printout, an "Output Complete" line indicates the transmission was
successful. If the "Output Complete" line is not present, ignore the data, since it may
be corrupted.
Real-Time Output
Real-time data is continuously sent to the data port on the back of the N-600x pulse
oximeter.
Patient data can be obtained through the data port by connecting the oximeter data
port to a PC or serial printer. When a real-time printout or display is being
transmitted to a printer or PC, a new line of data displays every second. Column
headings display or print after every 25 lines, or if one of the values in the column
heading changes. Readings display at one-second intervals.
Note:
If the data output stops transmitting, turn the power off and back on again or, if the oximeter
is connected to a PC, send an XON (Ctrl-q) command to reset the oximeter.