Modifying and Testing the Oximeter
10. Press the SRC-MAX MODULATION selection button. The SRC-MAX
MODULATION
in amplitude.
a. Active audio alarm.
b. Flashing %SpO
c. BPM indication between 57 and 63 inclusive.
d. Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual amplitude may
vary, but can be used as a reference for low pulse amplitude/low light patients.
11. Turn off the oximeter.
Light Level Test
1. With the oximeter turned off, connect the DOC-10 pulse oximetry cable to the
sensor port.
2. Connect the SRC-MAX tester to the other end of the DOC-10 cable.
3. Turn on the oximeter by pressing the ON/STANDBY button. Wait for the oximeter
to complete its self-check.
4. Select the PLETH view by pressing the following softkeys in sequence: SETUP,
VIEW, and then PLETH to view the pleth display.
a. Active audio alarm.
b. Flashing %SpO
c. BPM indication between 57 and 63 inclusive.
d. Pulse waveform of approximately 1/2-inch peak to peak (P-T-P) amplitude. Actual
amplitude may vary but will be a reference for low pulse amplitude/low light patients.
126
LED lights. The oximeter pulse amplitude waveform decreases
Figure 96. PLETH Display with SRC-MAX decrease in modulation
indication between 73 and 77 inclusive.
2
Figure 97. PLETH Display with SRC-MAX tester-generated waveform
indication between 73 and 77 inclusive.
2
10030881 Rev. B 12-2008
Service Manual