Operating the Oximeter
Note:
When the oximeter is in the FAST response mode, the oximeter may produce more SpO
pulse rate alarms than expected.
alarm behavior
4. Use the ADJUST UP or ADJUST DOWN keys below the
panel to select the desired response mode.
5. Press the EXIT softkey.
LIGHT Menu
1. OFF softkey—Use the OFF softkey to turn the backlight off entirely. This option
does not remain after power cycle, but returns to the factory default brightness.
Use the ADJUST UP or ADJUST DOWN keys below the ALARM SILENCE key on the
oximeter panel to brighten or dim the display backlight.
2. EXIT softkey—Exit to the Main Menu.

Adjusting the Factory Default Settings

Overview
WARNING
Audible alarms should not be silenced if patient safety could be compromised.
WARNING
Each time the pulse oximeter is used, check alarm limits to ensure that they are
appropriate for the patient being monitored.
WARNING
The SPD feature is for use with adults patients only. In Neonate Mode, the SPD feature
remains OFF.
Caution
Use of the SPD alarm feature does not change the need to set threshold limits appropriate to
the patient being monitored.
The OxiMax N-600x pulse oximeter is shipped with factory default settings. Factory
default settings are divided into two groups: adult and neonate. The oximeter lists
adult-pediatric alarm limit settings rather than neonate settings at power-up. Set the
oximeter's operating mode to adult-pediatric or neonatal using the LIMITS softkey.
This setting remains active until the oximeter is turned OFF. As a qualified service
technician, set institutional default settings, if they are different than the power-on
default settings, using Service Function Menus, page 130.
Use the softkeys to change alarm limits, displays, baud rates, time and date, and
trend data views. Some values cannot be saved as power-on default values.
The oximeter will not accept an %SpO
The oximeter will not accept an AUDIBLE ALARM OFF option as a power-on default.
72
The response mode, however, may impact the
.
Lower Alarm Limit of less than "85" as a power-on default.
2
10030881 Rev. B 12-2008
ALARM SILENCE
key on the oximeter
and
2
SPD
Service Manual