Oximeter Disassembly and Reassembly

Disassembling the Oximeter

Caution
Observe ESD (electrostatic discharge) precautions when disassembling and reassembling the
OxiMax N-600x pulse oximeter and when handling any of the components of the oximeter.
Note:
The battery charge procedure should be performed before oximeter repairs when possible.
To disassemble the oximeter:
1. Turn the oximeter off by pressing the ON/STANDBY button.
2. Disconnect the oximeter from the AC power source.
3. Set the oximeter upside down on a static-free work surface.
Caution
Ensure the work surface is clean and free of debris.
4. Using the phillips-head screwdriver, loosen and then remove the oximeter's four
corner screws.
5. Set aside all four screws for use during reassembly.
6. Carefully set the oximeter right-side up, grasping both halves to prevent
premature separation.