Repairing the Oximeter
Power Entry Module (PEM) Replacement
Removing the Power Entry Module
1. Turn the oximeter off by pressing the ON/STANDBY button.
2. Disconnect the oximeter from the AC power source and follow the steps outlined
in Disassembling the Oximeter, page 167.
3. Push the top of the Power Entry Module (PEM) in from the outside of the case, and lift up.
4. Use needle-nose pliers to disconnect the leads from the PEM as shown.
172
To Purchase, Visit
Avobus.com
Figure 128. Removing the Power Entry Module
1
Table 37. Removing the PEM Figure Callouts
No
1
N Connector
2
G Connector
3
L Connector
4
Equipotential Terminal (Ground)
10030881 Rev. B 12-2008