Error Codes - Nellcor OXIMAX N-600X Series Service Manual

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Troubleshooting
Message
READING
TRENDS...
REPLACE
SENSOR
RESET
DEFAULTS?
SAVE
DEFAULTS?
SENSOR
DISCONNECTED
SENSOR TYPE

Error Codes

When the OxiMax N-600x pulse oximeter detects an error condition, it displays "EEE"
followed by an error code of up to three digits.
144
Table 24. Common Prompts and Error Messages
Time-
Exit on
out in
Exit on
Alarm
secs
Alarm?
Silence?
None
Yes
Yes
None
No
Yes
10
No
Yes
10
No
Yes
None
No
Yes
5
No
No
Figure 120. Error Condition Screen, Battery Failure
10030881 Rev. B 12-2008
Displayed
When the oximeter
Completely retrieve sensor
needs more than
data or select ABORT.
four to six seconds to
retrieve trend data
from memory.
When the sensor or
Temporarily acknowledge by
cable wiring is
pressing the ALARM SILENCE
defective.
key. Replace sensor.
When attempting to
Leave default settings as they
reset oximeter to
are by selecting NO, or revert
factory default
back to the original factory
setting by pressing
default settings by selecting
the RESET softkey on
YES.
the PARAMS menu.
When attempting to
Confirm the new institutional
save current settings
default settings by selecting
as power-on defaults
YES, or leave default settings
by pressing SAVE in
as they are by selecting NO,
the PARAMS menu.
When the sensor is
Temporarily acknowledge by
disconnected from
pressing the ALARM SILENCE
the pulse oximetry
key. Connect the cable and
cable, the cable is not
sensor to the oximeter,
connected to the
reconnect the sensor, or
oximeter, or the cable
check all connections. If this
wiring is defective.
does not clear the condition,
replace the cable and/or
sensor.
First message
Automatically clears on
displayed when a
time-out.
sensor is connected
to the oximeter.
Resolution
Service Manual