Operating the Oximeter
2. SENSOR Submenu—The SENSOR menu is only available when using a single-
patient OxiMax pulse oximetry sensor. After connecting an OxiMax pulse
oximetry sensor, the bottom indicator on the left will blink for 60 seconds.
a. GRAPH softkey—Display events in reverse chronological order from top to bottom of
the display. View previous or next graphs if available. Return to the SENSOR Menu
using the BACK softkey.
b. TABLE softkey—Display events in graph form. Show previous or next tables if
available. PRINT table data or go BACK to the SENSOR Menu.
SETUP Menu
Use the SETUP menu to choose the preferred display screen, set temporary limits
and sensitivity, set oximeter time and date, select the preferred language, or select
the communication protocol, nurse call features, and response mode from the setup
menu. With a pulse oximetry sensor that retains the trend data, access the sensor
trend history data.
1. VIEW Menu—Select plethysmographic waveform, blip display, trend data display
or general care format (GCF) display view. Returns to default display after power
cycle.
a. PLETH softkey—Access plethysmographic (pleth) waveform display.
b. BLIP softkey—Access blip bar display.
c. TREND Menu—Access trend data display. Similar to MONITR trend view menu
option.
-
VIEW softkey—Display sensor trend data. Isolate oxygenation (SpO
pulse (PULSE Option) trend data or viewing them both together (DUAL Option).
-
ZOOM softkey—Access to the TIME Option, SCALE Option, AUTO Option, and
BACK Option if the trend view is set in DUAL, SpO
d. GCF softkey—Access the default general care format (GCF) display, which includes
the blip bar as well as SpO
SPD icons in a large, easy-to-view format.
2. SENSOR Menu—Access the sensor trend history menu. This menu is only
available if you have a pulse oximetry sensor type that retains the trend data in a
chip on the pulse oximetry sensor.
a. DATA softkey—Access the sensor trend history data. The unit identifies which type of
pulse oximetry sensor it is, as well as type of available data.
b. MSG softkey—Determine of messaging is enabled or disabled (Yes or No) and review
sensor events (Yes or No).
3. CLOCK Menu—Access to set the clock, both date and time.
WARNING
The pulse oximetry sensor extrapolates from the date and time provided by the Nellcor
OxiMax N-600x pulse oximeter when recording the sensor event record to the sensor.
The accuracy of the date/time is determined by the date/time setting of the pulse
oximeter. Set the pulse oximeter date and time to the correct value before connecting