Reassembling the Oximeter
To reassemble the oximeter:
1. Connect the oximeter's power supply harness to J200 (3) on the Main Board PCB.
2. Place the oximeter's top enclosure over the bottom case, being careful to align the
Display PCB, Power Entry Module, and the fan with the slots in the oximeter casing.
Caution
Do not pinch any wires between the cases when reassembling the oximeter. Before installing
the screws, inspect the entire edge of the union for even seating.
3. Grasp both halves, gently rotating it upside-down.
4. Place on a static-free work surface.
5. Replace the four corner screws set aside during disassembly in the bottom cover
of the unit.
6. Gently tighten the screws that hold the case halves together (10-in lbs.
maximum).
Caution
Overtightening could strip out the screw holes in the top enclosure, rendering it unusable.
Service Manual
To Purchase, Visit
Avobus.com
10030881 Rev. B 12-2008
or call
1-800-674-3655
Oximeter Disassembly and Reassembly
169