14 Repairing the Oximeter

Overview

WARNING
Only qualified service personnel should open the oximeter housing, remove and
replace components, or make adjustments. If your medical facility does not have a
qualified service technician, please contact Nellcor's Technical Services or your local
Nellcor representative.
WARNING
Before attempting to open or disassemble the OxiMax N-600x pulse oximeter,
disconnect the power cord to avoid possible injury.
WARNING
No user serviceable parts inside.
Caution
Observe ESD (electrostatic discharge) precautions when working within the unit.
OxiMax N-600x™ pulse oximeter major component parts include all PCBs and
subassemblies.
PCBs
Battery
Cables
Chassis enclosures
Collect the following tools prior to any disassembly of the oximeter.
Small- and medium-sized phillips-head screwdriver
Small flat-blade screwdriver
Needle nose pliers or 1/4-inch socket with driver
10mm wrench
Torque wrench, 10-in lbs. (1.13 Newton-meters)
Note:
The battery charge procedure should be performed before oximeter repairs when possible.
Only a qualified service technician may disassemble the oximeter to its major
component parts. The supported replacement level for the OxiMax N-600x pulse
oximeter is to the printed circuit board (PCB) and major subassembly level. After
isolating the problem to a suspected PCB, follow the procedures in Disassembling the
Service Manual