Alarm Speaker Replacement

Removing the Alarm Speaker

O
M
N-600x Service Manual
XI
AX
6. Insert the J200 cable into the Main Board PCB until it locks into place.
7. Connect the speaker cable to J9 on the Main Board PCB.
8. Connect the CCFL wire harness with two white wires to J4 on the Main Board
PCB.
9. Connect the Display PCB ribbon cable to J7 on the Main Board PCB. Install
the clip over the J7 connector.
10. Follow the steps outlined in Assembling the Monitor on page 88.
To remove the alarm speaker:
1. Turn the monitor off by pressing the ON/STANDBY button.
2. Disconnect the monitor from the AC power source and follow the steps
outlined in Disassembling the Monitor on page 85.
3. Disconnect the speaker wire harness from J9 on the Main Board PCB shown
in Alarm Speaker on page 100.
Disassembling the Monitor