Primary Speaker Failure - Nellcor OxiMax N-600X Operator's Manual

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

Primary Speaker Failure

N-600x Operator's Manual
The N-600x may detect a failure of the primary speaker and sound a
high-pitched, slow-pulsing piezo tone. A primary speaker failure
message displays as shown below.
PRIMARY SPKR FAILURE:
NOTIFY SVC PERSONNEL.
PRESS HELP
HELP
1. Press HELP to continue. The following message displays.
HELP SPEAKER FAILURE
Note: Once this
monitor is powered
off, it cannot be
powered on again.
2. Press BACK to display the speaker failure message again. The
message cannot be cleared.
3. Press the ALARM SILENCE button to silence the slow-pulsing
piezo tone.
Note: Once the monitor is silenced, the N-600x sounds a piezo tone
every three minutes as a reminder of the primary speaker
failure condition. The N-600x also sounds the piezo tone to
annunciate low, medium and high priority alarms during this
time. If an N-600x monitor is reporting a primary speaker
failure is powered off, it cannot be powered on again.