Disassembly Guide; Introduction; Tools Required; Replacement Level Supported - Nellcor OxiMax N-65 Service Manual

Handheld
Hide thumbs Also See for OxiMax N-65:
Table of Contents

Advertisement

D i s a s s e m b l y G u i d e

Introduction

Tools Required

Replacement Level Supported

O
M
N-65 Service Manual
XI
AX
The O
M
N-65 can be disassembled into all major component parts, including:
XI
AX
Front case with front panel keypad
Rear bottom
User Interface PCB
SpO
PCB
2
Battery compartment door
Note: Some spare parts have a business reply card attached. Please fill out and
return the business reply card.
Caution: Observe ESD (electrostatic discharge) precautions when
disassembling and reassembling the monitor and when handling any
components of it.
The tools required to disassemble and reassemble the monitor are:
Number one (medium) Phillips screwdriver
Torque driver, four inch/pounds (10 cm/kg)
The replacement level supported for the monitor is at the printed circuit board
(PCB) and major subassembly level. When a PCB problem is identified, follow
the procedures in Disassembly Guide on page 47 to replace the PCB. Check to
ensure the symptom disappears and that the monitor passes all performance tests.
If the symptom persists, swap the replacement PCB with the malfunctioning PCB
(the original PCB that was installed when you started troubleshooting) and
continue troubleshooting.
47

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents