DD+DIS170.11E
Upon digitizer installation, it must be ensured that either the mains connector or an all-
pole circuit breaker for the installation on site is located close to the machine and easily
accessible.
6.4
Network Connection in the Installation Room
The network connection in the installation room has to comply with IEC 60950.
6.5
Mains cable
Following cables are part of delivery:
Euro*
Version:
*
Exceptions:
Denmark,
Ireland, Italy
Malta, Cyprus,
UK and
Switzerland
US/CAN-
Version:
NOTE:
The digitizer is delivered with two power cables for US and Europe. If a different cable
is needed, organize it locally but keep attention to country specific requirements.
Example:
Power cable to be organized locally for US/CAN, if a connection to a 240 V-outlet
instead of the standard 120 V-outlet is needed:
NEMA 6-15P - SJT, 3 x AWG 18 with IEC C13 connector (female); Hospital grade (HG)
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
CEE7/7 - H05VV - F3G1 - 3 x 1 mm² with
IEC C13 plug (female);
cable length 3 m
NEMA 5-15P - SJT,
3 x AWG 16 with
IEC C13 plug (female); Cable length 3 m,
Hospital grade (HG)
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 11 / Page 16 of 24
Type 5151 / 100 / 110 / 200 / 210
Installation Planning
IEC C13
Figure 8
NEMA 5-15P
Figure 9
Agfa Company Confidential
1
2
3
CEE7/7
5175_C11_001.cdr
IEC C13
5175_C11_002.cdr