Endurance Run Cycle - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
2.5.4

Endurance Run Cycle

The menu "Endurance Run Cycle" allows to run up to 300 cassette cycles with or
Purpose
without scanning and erasure:
This can be helpful when investigating sporadic image plate handling failures.
(1)
Select service menu: <Diagnose
Usage
cycle>: The Error Viewer window
opens.
(2)
Click: <Confirm>
(3)
Enter the number of cycles
(4)
Select <NO> for Scanning (i.e. laser
beam switched off during scan) and /
or Erasure if required.
(5)
Put a cassette into the digitizer.
(6)
Click: <Start>
(7)
Click <Stop> to interrupt before the
number of adjusted cycles is
completed.
IMPORTANT:
The Error Viewer and related confirm / cancel window only pops up at the
NX workstation. This means, that in any case, entering the Endurance Run
window needs to be confirmed at the NX workstation.
Only leave the Endurance Run window by the CANCEL button. Leaving it via
browser back button requires a reboot of the digitizer to allow further service
activities.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 23 of 31
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Tools and auxiliary Means
Figure 19
Figure 20
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents