Restore - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
2.7.2

Restore

The menu "Restore" is required to restore a certain configuration of the digitizer.
Purpose
(1)
Select service menu: <Restore>
Usage
(2)
Click: <Read from USB> or <Read from PC>
(3)
Select the backup file. It is a file with following syntax:
BAC_CR10-5151_<Serial number>DDMMMYYYY_HH-MM-SS.ZIP
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 29 of 31
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Tools and auxiliary Means
Figure 26
Agfa Company Confidential