Software Release Arc_1202; Solved Software Issues In Software Arc_1202 - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
3

Software Release ARC_1202

Release Date
Digitizer SNs
Availability
Prerequisites
Reference Document
* On the Agfa HealthCare Library only the latest software release can be downloaded.
3.1

Solved Software Issues in Software ARC_1202

In software ARC_1202 the Galvo adjustment time is adapted: This prevents error
10852 (Adjust Galvo failed) which may appear at a digitizer fully warmed up and
operated at higher ambient air temperature.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
04-2013
07/2012
Software ARC_1202 is installed in digitizers with
SN ≥ 1215 and additionally in 1206, 1211 and SN 1212
ex factory.
The NX workstation connected to the digitizer requires
software ≥ NX 2.0.8700 / 3.0.8700.
For additional system component compatibility
information refer to following documents:
DD+DIS184.12E, Service Bulletin No. 03,
Software Update ARC_1202 available to solve Galvo
Error 10852
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.8 / Page 7 of 13
Type 5151 / 100 / 110 / 200 / 210
Software Releases and Patches
Agfa HealthCare Library*
USB flash drive, delivered with the digitizer
Solution Structure Data Document, Document ID*
30781759
(Intranet Link
CR/DR Interoperability Matrix, Document ID*
31333326
(Intranet Link
(Intranet Link
Repair and Service
/
Extranet
Link)
/
Extranet
Link)
/
Extranet
Link)
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents