Software Release ARC_1304

Release Date
Digitizer SNs
Availability
Prerequisites
Reference Document
* On the Agfa HealthCare Library only the latest software release can be downloaded.
5.1

Solved Software Issues in Software ARC_1304

Software ARC_1304 solves some software issues, that appeared in production only:
During boot-up the target high voltage for the Photomultiplier is not reached. Boot-up
stops with error ARC2202 (10753). In software ARC_1304 the adjustment for the target
high voltage is optimized to prevent the issue (HQ_1210190006).
When the image plate is slipping on the scan drum before scanning starts, the slow
scan motor decelerates already before the last few scan lines were scanned. This may
lead to a decreased density at the last few scan lines (< 1 mm scan length)
(HQ_121008000).
In software ARC_1304 scanning is extended by 3 mm to have enough safety margin.
Warning messages about scan length deviations are issued in the Error Viewer,
despite no image distortions are available. In software ARC_1304 the mechanism to
display the warning message is less sensitive.
Sporadically the Begin of Scan signal is detected too early, due to a timing problem. In
this case error ARC4200 (4254) is displayed. After a reboot the image is usually
scanned without error message. In software ARC_1304 the timing problem has been
corrected.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
04-2013
10/2012
Software ARC_1304 is installed in digitizers with SN ≥
1551 and additionally in SN 1377, 1381, 1405, 1424,
1428, 1482, 1495, 1516 to 1523, 1525 to 1528, 1533,
1534, 1537 to 1549 ex factory.
The NX workstation connected to the digitizer requires
software ≥ NX 2.0.8700 / 3.0.8700.
For additional system component compatibility
information refer to following documents:
DD+DIS189.12E, Service Bulletin No. 06, Software
Update ARC_1304 available
Link)
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.8 / Page 10 of 13
Type 5151 / 100 / 110 / 200 / 210
Software Releases and Patches
Agfa HealthCare Library*
USB flash drive, delivered with the digitizer
Solution Structure Data Document, Document ID*
30781759
(Intranet Link
CR/DR Interoperability Matrix, Document ID*
31333326
(Intranet Link
Repair and Service
/
Extranet
Link)
/
Extranet
Link)
(Intranet Link
/
Extranet
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents