Select service menu: <Clear Image
Usage
Queue>
(2)
Select: <Clear Image Queue>.
(3)
Click: <OK>
NOTE:
This menu is not available for software < ARC_1404.
2.6
Maintain & Calibrate
2.6.1

Shading Calibration

The menu "Shading Calibration" provides:
Purpose
Functionality to compensate image inhomogeneities caused by certain properties
of the scanning system of the digitizer (Button "Calibrate").
Possibility to view the current calibration curve (Link "Draw calibration line") for
troubleshooting.
Possibility to download the current calibration curve to PC (Link "Download
calibration line") for troubleshooting. Note, that the calibration curve is also part of
a complete backup.
A shading calibration has to be performed after replacement of following spare parts:
Optic Module
PMT with Light Collector
To perform a shading calibration:
Usage
(1)
Select service menu: <Shading Calibration>
(2)
Follow the instructions on the screen and in CR 10-X Service Manual,
chapter 3.6, Adjustments and Calibrations.
To view the current calibration curve for troubleshooting purpose:
(1)
Select link: <Draw calibration line>
To copy the current calibration curve to PC for troubleshooting purpose:
(1)
Select link: <Download calibration line>
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 26 of 31
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Tools and auxiliary Means
Figure 23
Agfa Company Confidential